C.R. BARD, INC. (BASD) -3006260740 APHERESIS PORT; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number UNK APHERESIS PORT ACCESS SYS |
Device Problems
Blocked Connection (2888); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Scar Tissue (2060)
|
Event Date 01/14/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
As the lot number for the device was not provided, a review of the device history records could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.
|
|
Event Description
|
It was reported that sometime post port placement procedure, the patient had scar tissue formation around the port.It was further reported that the device was unable to access due to the scar tissue formation.Current status of the patient is unknown.
|
|
Event Description
|
It was reported that sometime post port placement procedure, the patient had scar tissue formation around the port.It was further reported that the device was unable to access due to the scar tissue formation.Current status of the patient is unknown.
|
|
Manufacturer Narrative
|
H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported scar tissue formation around the port issue as no objective evidence was provided for review.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Search Alerts/Recalls
|
|
|