MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. DERMATOME AN HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. DERMATOME AN HANDPIECE Back to Search Results
Model Number N/A
Device Problem Structural Problem (2506)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2022
Event Type  malfunction  
Manufacturer Narrative
This complaint has been recorded under (b)(4).Once investigation is completed a supplemental/final report will be submitted.
 
Event Description
It was reported that prior to surgery it was noted the device was missing a screw.No harm or impact to graft were noted.Another device was used to complete the procedure.No other adverse event was reported as it relates to this event.
 
Event Description
No additional information available.
 
Manufacturer Narrative
This complaint has been recorded under (b)(4).Review of the most recent repair record determined the width plate screw was missing and was replaced which resolved the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DERMATOME AN HANDPIECE
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key13506345
MDR Text Key285647308
Report Number0001526350-2022-00123
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024464834
UDI-Public(01)00889024464834(11)190314(10)64324727
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number88710100
Device Lot Number64324727
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Date Manufacturer Received03/31/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/14/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-