Model Number N/A |
Device Problems
Mechanical Problem (1384); Structural Problem (2506)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/03/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the item was sent in for preventative maintenance and it was found that the device was missing screws, hinges, and had a defective motor.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
|
|
Manufacturer Narrative
|
This complaint has been submitted under (b)(4).Review of the most recent repair record determined the unit was found to have an unstable motor speed, a missing hinge gasket, and the width plate screws were worn.The motor, hinge gasket, and width plate screws were replaced and resolved the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional information available.
|
|