Model Number 10-80-00
Device Problem Decreased Pump Speed (1500)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2022
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could not reproduce the reported issue.The pump was found to be working within specifications.No parts related to the reported event were replaced.Since no failure on the device could be identified, the most likely root cause of the reported issue is over-occlusion of tubings set by the user.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report of a s5 roller pump that did not run at constant speed when at low speed.No additional detail was provided.There was no report of patient injury.
 
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could not reproduce the reported issue.The pump was found to be working within specifications.No parts related to the reported event were replaced.Since no failure on the device could be identified, the most likely root cause of the reported issue is over-occlusion of tubings set by the user.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report of a s5 roller pump that did not run at constant speed when at low speed.No additional detail was provided.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key13535164
MDR Text Key285689201
Report Number9611109-2022-00068
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 02/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/22/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse