• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MOOG MEDICAL DEVICES GROUP ENTERALITE INFINITY ENTERAL FEEDING PUMP; ENTERAL INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MOOG MEDICAL DEVICES GROUP ENTERALITE INFINITY ENTERAL FEEDING PUMP; ENTERAL INFUSION PUMP Back to Search Results
Model Number INFKIT2
Device Problem Fire (1245)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/21/2022
Event Type  malfunction  
Manufacturer Narrative
The pump was returned to mmdg for evaluation.A dhr review was completed and found no non conformances.When the device was returned to mmdg for investigation, the pump operated as expected.There was also no indication that the pump had experienced any kind of fire.This report is being filed because all complaints of this type are considered reportable by mmdg.
 
Event Description
The initial reporter stated that the pump "sparked".Mmdg made multiple attempts to follow up with the initial reporter and obtain additional information, but those attempts were unsuccessful.[complaint-(b)(4)].
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERALITE INFINITY ENTERAL FEEDING PUMP
Type of Device
ENTERAL INFUSION PUMP
Manufacturer (Section D)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city UT 84123
Manufacturer (Section G)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city UT 84123
Manufacturer Contact
kristin egelhoff
4314 zevex park lane
salt lake city, UT 84123
8012641001
MDR Report Key13538544
MDR Text Key285664253
Report Number1722139-2022-00182
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031199
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberINFKIT2
Device Catalogue NumberINFKIT2
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/15/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-