• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES CB030, 5MMX35CM EPIX LAPA SCISSOR, 10/BX; LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES CB030, 5MMX35CM EPIX LAPA SCISSOR, 10/BX; LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES) Back to Search Results
Model Number CB030
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2021
Event Type  malfunction  
Manufacturer Narrative
The event device is anticipated to return.A follow-up report will be provided upon completion of the investigation.
 
Event Description
Name of procedure being performed: na (incoming inspection).Detailed description of event: [facility] incoming inspection po# 154p055.This is a complaint from aomori olympus evaluation team.The lot mentioned above was found to have defective units per our incoming inspection.Delivery date: december 17, 2021.We found defects as the result of incoming inspection of the products.Please refer to the attachment material for the quantity which returns the products.Hairlike object in pouch - cb030 lot #1431233 (1 ea.).Patient status: na (no patient involvement).Type of intervention: na (incoming inspection).
 
Manufacturer Narrative
The event unit was returned to applied medical for evaluation.Visual inspection confirmed the presence of hair inside the sterile unopened pouch.Based on the condition of the returned unit and the description of the event, the hair likely originated from an operator during the manufacturing process.
 
Event Description
Name of procedure being performed: na (incoming inspection).Detailed description of event: [facility] incoming inspection po# (b)(4).This is a complaint from [facility] evaluation team.The lot mentioned above was found to have defective units per our incoming inspection.Delivery date: december 17, 2021.Please refer to the attachment file.We found defects as the result of incoming inspection of the products.Please refer to the attachment material for the quantity which returns the products.Hairlike object in pouch - cb030 lot #1431233 (1 ea.).Patient status: na (no patient involvement).Type of intervention: na (incoming inspection).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CB030, 5MMX35CM EPIX LAPA SCISSOR, 10/BX
Type of Device
LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES)
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
aaron fulcher
22872 avenida empresa
rancho santa margarita, CA 92688
9497135765
MDR Report Key13539018
MDR Text Key287252847
Report Number2027111-2022-00371
Device Sequence Number1
Product Code HET
UDI-Device Identifier00607915110123
UDI-Public(01)00607915110123(17)241003(30)01(10)1431233
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K062169
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberCB030
Device Catalogue Number101419701
Device Lot Number1431233
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Date Manufacturer Received01/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/04/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-