MEDOS INTERNATIONAL SÃ RL CH EXPEDIUM SPINE SYSTEM SINGLE INNER SET SCREW 5.5; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
|
Back to Search Results |
|
Model Number 179702000 |
Device Problem
Migration (4003)
|
Patient Problems
Purulent Discharge (1812); Non-union Bone Fracture (2369)
|
Event Type
Injury
|
Manufacturer Narrative
|
Additional narrative: complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on an unknown date, the patient underwent a removal procedure due to severe pus in the lesion.During the removal procedure the setscrew chipped.The surgeon removed two fragments from the patient¿s lesion.No more fragments were left in the patient.The procedure was completed without surgical delay.The patient received the implants on (b)(6) 2015 and the bone fusion had not been attained at the time of the removal surgery.This report involves one (1) expedium spine system single inner set screw 5.5.This is report 1 of 5 for (b)(4).
|
|
Manufacturer Narrative
|
Additional narrative: complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on an unknown date, the patient underwent a removal procedure due to severe pus in the lesion.During the removal procedure the setscrew chipped.The surgeon removed two fragments from the patient¿s lesion.No more fragments were left in the patient.The procedure was completed without surgical delay.The patient received the implants on (b)(6) 2015 and the bone fusion had not been attained at the time of the removal surgery.This report involves one (1) expedium spine system single inner set screw 5.5.This is report 1 of 5 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the product was returned to us customer quality (cq) for evaluation.The us cq team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the single-inner setscrew had broken threads.The broken thread pieces were also returned.No other issues were identified.The device migration complaint condition reported cannot be confirmed as no radiographic images were provided.The dimensional inspection was not performed due to the post manufacturing damage.The observed condition of the single-inner setscrew was consistent with a random component failure that may have been caused by exposure to excessive/unintended forces.A functional test could not be performed as the mating devices were not returned.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the single-inner setscrew was observed to have broken threads.While no definitive root cause could be determined from the available information, it is probable that the threads of the single-inner setscrew were broken due to the exposure to excessive/unintended forces.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|