MEDOS INTERNATIONAL SàRL CH EXPEDIUM SPINE SYSTEM SINGLE INNER SET SCREW 5.5; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 179702000
Device Problem Migration (4003)
Patient Problems Purulent Discharge (1812); Non-union Bone Fracture (2369)
Event Type  Injury  
Manufacturer Narrative
Additional narrative: complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on an unknown date, the patient underwent a removal procedure due to severe pus in the lesion.During the removal procedure the setscrew chipped.The surgeon removed two fragments from the patient¿s lesion.No more fragments were left in the patient.The procedure was completed without surgical delay.The patient received the implants on (b)(6) 2015 and the bone fusion had not been attained at the time of the removal surgery.This report involves one (1) expedium spine system single inner set screw 5.5.This is report 1 of 5 for (b)(4).
 
Manufacturer Narrative
Additional narrative: complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on an unknown date, the patient underwent a removal procedure due to severe pus in the lesion.During the removal procedure the setscrew chipped.The surgeon removed two fragments from the patient¿s lesion.No more fragments were left in the patient.The procedure was completed without surgical delay.The patient received the implants on (b)(6) 2015 and the bone fusion had not been attained at the time of the removal surgery.This report involves one (1) expedium spine system single inner set screw 5.5.This is report 1 of 5 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the product was returned to us customer quality (cq) for evaluation.The us cq team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the single-inner setscrew had broken threads.The broken thread pieces were also returned.No other issues were identified.The device migration complaint condition reported cannot be confirmed as no radiographic images were provided.The dimensional inspection was not performed due to the post manufacturing damage.The observed condition of the single-inner setscrew was consistent with a random component failure that may have been caused by exposure to excessive/unintended forces.A functional test could not be performed as the mating devices were not returned.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the single-inner setscrew was observed to have broken threads.While no definitive root cause could be determined from the available information, it is probable that the threads of the single-inner setscrew were broken due to the exposure to excessive/unintended forces.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPEDIUM SPINE SYSTEM SINGLE INNER SET SCREW 5.5
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kara ditty-bovard
chemin-blanc 38
le locle 02400
SZ   02400
6103142063
MDR Report Key13543225
MDR Text Key285693031
Report Number1526439-2022-00223
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034064307
UDI-Public10705034064307
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K160904
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number179702000
Device Catalogue Number179702000
Device Lot NumberATDD08
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/12/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
5.5 TI CORT FIX 7X40MM; 5.5 TI CORT FIX 7X45MM; UNK - RODS: EXPEDIUM; VIPER TI SAI POLY 9X80MM
Patient Outcome(s) Required Intervention;
-
-