COLTENE/WHALEDENT INC. HYGENIC; DENTAL DAM CLAMP Back to Search Results
Catalog Number H01049
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2022
Event Type  Injury  
Event Description
On (b)(6) 2022 the customer stated the following: clamp breaks in the patient's mouth under rubber dam so there is no debris in the mouth.It is a fracture at the level of the arch of the clamp.The metal deforms and breaks after one or two uses.These incidents are numerous for b2 a little less b3 almost never b1.The metal does not resist the clamp pliers which open the clamp before positioning on the tooth.And it breaks during the treatment a little after putting it on.At this time, the field also leaves.The session must be stopped.Take another clamp and start over.The additional information provided states that there was no patient or user injury.The dental dam came off the tooth when the clamp broke, and all pieces of the clamp were retrieved.
 
Manufacturer Narrative
This complaint is for material h01049 b 2 hyg tissue retractor.No batch information was provided so an inventory analysis cannot be performed.The clamp was returned for investigation.This clamp has a date code of pd which means it was manufactured in 2019 july/aug.The clamp was examined under a microscope.The surface of the clamp is pitted and there are a couple of small rust spots.Additionally, the break surface has black discoloration and a couple of small rust spots.The customer's cleaning process including equipment and solutions is unknown.However, the condition of the break surface indicates that the customer likely used a solution in their cleaning process which was able to corrosively penetrate the material.The black coloration indicates that the ph of the cleaning solution was too low (acidic).Use of a corrosive cleaning solution can cause the degradation of the material of the clamp.It is most likely that the degradation of the material of the clamp caused the breakage.Clamps undergo an amount of stress when they are expanded under normal use.Degradation of the material causes them to be weaker and unable to withstand the force of expansion.To prevent degradation of the material, clamps need to be cleaned with a ph neutral solution and to be rinsed properly with water after cleaning.De-ionized is preferable for rinsing as tap water may contain harsh alkaline chemicals (chlorides) added to kill bacteria.Root cause: break caused by material degradation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYGENIC
Type of Device
DENTAL DAM CLAMP
Manufacturer (Section D)
COLTENE/WHALEDENT INC.
235 ascot parkway
cuyahoga falls OH 44223
Manufacturer (Section G)
COLTENE/WHALEDENT INC.
235 ascot parkway
cuyahoga falls OH 44223
Manufacturer Contact
tricia cregger
235 ascot parkway
cuyahoga falls, OH 44223
MDR Report Key13549298
MDR Text Key286018586
Report Number2416455-2022-00002
Device Sequence Number1
Product Code EEF
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 04/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/17/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberH01049
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/20/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-