Model Number N/A |
Device Problems
Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923); Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Failure of Implant (1924); Pain (1994); Joint Dislocation (2374)
|
Event Date 01/21/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Medical products: item#: 110030777, cr 40mm glenosphere +3mm cocr; lot#: 64328699.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported patient underwent a revision procedure approximately seven months post implantation due to pain and disassociation of the glenosphere from the baseplate of their left shoulder.A new augmented baseplate, glenosphere/tray/poly were implanted.Attempts to obtain additional information have been made; however, no more is available.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.Visual examination of the provided pictures identified all explanted devices, wear can be seen the bearing.However, as product was not returned visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: revision surgery (b)(6) 2022.The patient presents with acute onset shoulder pain due to displacement of the left shoulder glenosphere when working on his truck.The displaced glenosphere was identified and removed.The poly liner was removed as there was significant scuffing noted.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that the patient was revised due to injuring themselves while working on their truck.Injury from accident caused the displacement of the glenosphere.
|
|
Search Alerts/Recalls
|