• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. COMP RVRS 25MM BSPLT HA+ADPTR; SHOULDER PROSTHESIS, REVERSE CONFIGURATION/EXTREMITIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. COMP RVRS 25MM BSPLT HA+ADPTR; SHOULDER PROSTHESIS, REVERSE CONFIGURATION/EXTREMITIES Back to Search Results
Model Number N/A
Device Problems Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923); Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Failure of Implant (1924); Pain (1994); Joint Dislocation (2374)
Event Date 01/21/2022
Event Type  Injury  
Manufacturer Narrative
(b)(4).Medical products: item#: 110030777, cr 40mm glenosphere +3mm cocr; lot#: 64328699.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported patient underwent a revision procedure approximately seven months post implantation due to pain and disassociation of the glenosphere from the baseplate of their left shoulder.A new augmented baseplate, glenosphere/tray/poly were implanted.Attempts to obtain additional information have been made; however, no more is available.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Visual examination of the provided pictures identified all explanted devices, wear can be seen the bearing.However, as product was not returned visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: revision surgery (b)(6) 2022.The patient presents with acute onset shoulder pain due to displacement of the left shoulder glenosphere when working on his truck.The displaced glenosphere was identified and removed.The poly liner was removed as there was significant scuffing noted.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the patient was revised due to injuring themselves while working on their truck.Injury from accident caused the displacement of the glenosphere.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP RVRS 25MM BSPLT HA+ADPTR
Type of Device
SHOULDER PROSTHESIS, REVERSE CONFIGURATION/EXTREMITIES
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13560503
MDR Text Key285829501
Report Number0001825034-2022-00371
Device Sequence Number1
Product Code KWS
UDI-Device Identifier00880304532465
UDI-Public(01)00880304532465(17)301210(10)633840
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193373
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/18/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number010000589
Device Lot Number633840
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/10/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age69 YR
Patient SexMale
Patient Weight88 KG
-
-