Model Number HL-90 |
Device Problems
Fire (1245); Device Emits Odor (1425)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/12/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to smiths medical.The root cause is main board, ac transformer.A manufacturing dhr review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.
|
|
Event Description
|
It was reported that,yesterday, operating room staff reported a burning smell coming from this unit.I opened the unit to find out that mainboard has been burnt just above f2.No patient injury was reported.
|
|