Model Number 7-5105 |
Device Problem
Moisture Damage (1405)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 01/06/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Moisture causing them to get stuck so, they would break apart or must be thrown out.The adapter is a component of a ventilator circuit and the incident occurred during use.Base on the reported information the criteria for reporting an adverse event has been met.
|
|
Event Description
|
Moisture causing them to get stuck, so they would break apart or have to be thrown out.
|
|
Event Description
|
Moisture causing them to get stuck so they would break apart or have to be thrown out.
|
|
Manufacturer Narrative
|
Moisture causing them to get stuck so they would break apart or must be thrown out.The adapter is a component of a ventilator circuit and the incident occurred during use.Base on the reported information the criteria for reporting an adverse event has been met.Complaint confirmed via testing of lot# 010422u26.22mm od side of adaptor is not iso spec, but actually 0.875" as noted on drawing 24011.Ncmr-02278 has been opened to address molding issues.Customer informed of nonconformance.Issue has been sent for carb review due to loss of intended function from being out of tolerance.
|
|