A field service engineering (fse) was at the customer's site to address reported event.The customer reported out of range high fsh qc result to fse.Fse confirmed the complaint by reviewing the error logs, but was not able to reproduce error.While troubleshooting, fse found the bush wash syringe was worn out and causing leaks.Fse replaced the bush wash syringe, performed a decontamination on the analyzer, calibrated and ran qc without error and within acceptable range.No further action required by field service.The aia-900 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were three (3) similar complaints identified during the searched period, which includes this event.The st aia-pack fsh analyte application manual states the following: evaluation of results quality control in order to monitor and evaluate the precision of the analytical performance, it is recommended that commercially available control samples be assayed daily.The minimum recommendations for the frequency of running internal control material are: after calibration, two levels of controls are run in order to accept the calibration curve.The two levels of controls are also repeated after calibration when certain service procedures are performed (e.G.Temperature adjustment, sampling mechanism changes, maintenance of the wash probe or detector lamp adjustment or change).After daily maintenance, at least two levels of the control should be run in order to verify the overall performance of the tosoh aia system analyzers.If one or more control sample value(s) is out of the acceptable range, it will be necessary to investigate the validity of the calibration curve before reporting patient results.Standard laboratory procedures should be followed in accordance with the regulatory agency.The most probable cause of the reported event is due to worn wash syringe.
|