Model Number N/A |
Device Problems
Incorrect Measurement (1383); Failure to Cut (2587)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that the depth measurement was inaccurate during pre-surgery.There was no harm or delay.No adverse events were reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once an evaluation of this device is completed, a follow-up/final report will be submitted.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once an evaluation of this device is completed, a follow-up/final report will be submitted.
|
|
Event Description
|
It was reported that the depth measurement was inaccurate during pre-surgery.There was no harm or delay.No adverse events were reported as a result of this malfunction.
|
|
Event Description
|
No additional event information.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the calibration was out at the 0 and 4 reading.The vespel and semi-circle bearings were replaced to aid in calibration and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|