The mayfield composite series skull clamp (a3059) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - evaluation found no device deficiencies that would have contributed to the reported complaint.The unit passed all specific functional testing requirements, except for the lock having rotational movement when the unit is not under pressure.However, this would not have caused a slippage.General maintenance and cleaning was performed.Root cause - evaluation found no device deficiencies that would have contributed to the reported complaint.Probable root cause of the reported incident is improper placement of the skull clamp on the patient or improper support of the patient's head.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
A facility reported that during "left parietal craniotomy for tumor biopsy" procedure, they could not advance the torque screw on the mayfield composite series skull clamp (a3059) pass 40pp.They released the patient's head, and it slipped resulting in an approximately.75¿ laceration on the forehead.Laceration was treated with stitches, and the patient was reported as stable.A different skull clamp was used without incident.No surgical delay has been reported.
|