Catalog Number 03666794001 |
Device Problems
Display Difficult to Read (1181); Display or Visual Feedback Problem (1184)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/03/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The meter was requested for investigation.
|
|
Event Description
|
The initial reporter stated there was an issue with the display of a coaguchek xs meter.The meter was powered on and had segments missing from the results field of the display.A display check of the meter was performed and only the tops of the "888" were seen in the results field.When checking the meter memory, a result of 19.1 seconds (41 % quick, 1.6 inr) was observed.
|
|
Manufacturer Narrative
|
The meter was returned for investigation.A display test showed no faded or missing segments.When displaying measurement results, some segments are displayed with poor contrast.The circuit board was tested for damage or contamination.The battery compartment and circuit board showed contamination by liquid (leaked battery) which has penetrated/corroded the solder contacts.The contamination also affected the conductive rubber contacts, causing the low contrast display.The contamination issue would be related to a handling error.
|
|