• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX BLUE LINE ULTRA SUCTIONAID; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX BLUE LINE ULTRA SUCTIONAID; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Model Number 101/870/075CZ
Device Problem Detachment of Device or Device Component (2907)
Patient Problems Fever (1858); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 01/01/2022
Event Type  Injury  
Event Description
It was reported that no anomaly was found in the product during the pre-use air leak check.However, seven (7) days after the customer started using it, he noticed the pilot balloon was detached from the inflation line.The patients had spontaneous breathing.
 
Manufacturer Narrative
Other, other text: one unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.The problem source is design.Similar customer complaints have been recently received therefore internal non-conformity report (b)(4) was created with purpose to evaluate this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BLUE LINE ULTRA SUCTIONAID
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
7-15-26 fukushima,
minneapolis, MN 55442
MDR Report Key13635392
MDR Text Key286342486
Report Number3012307300-2022-04222
Device Sequence Number1
Product Code BTO
UDI-Device Identifier15019517076219
UDI-Public15019517076219
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K173384
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number101/870/075CZ
Device Catalogue Number101/870/075CZ
Device Lot Number4138176
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/30/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/05/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-