• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 CATHERA; CATHETER, CONTINUOUS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 CATHERA; CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number FG11150-0615-2X
Device Problems Compatibility Problem (2960); Material Integrity Problem (2978); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2022
Event Type  malfunction  
Manufacturer Narrative
Healthcare professional information is not provided or reported due to country privacy laws.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a phenom catheter that had resistance and was occluded.The patient was undergoing a coil embolization to treat a carotid cavernous fistula (ccf).Vessel tortuosity was moderate.It was reported that the phenom catheter was prepared and flushed with saline per ifu.Several coils were successfully delivered by the catheter.Then, a coil was stuck during insertion.It was thought that the catheter was occluded.The physician thought the internal coating of the catheter might have peeled off and stuck causing the occlusion/resistance.The phenom catheter was removed and replaced with another manufacturer's device to complete the procedure.There was no harm or injury to the patient.
 
Manufacturer Narrative
H3.Product analysis: equipment used: vis (m-78210), 203cm ruler (m-83361), in-house mandrel as found condition (condition of returned device): the phenom-17 catheter was returned for analysis within a shipping box and within a resealable biohazard pouch.The unknown coil used in the event was not returned.Visual inspection/damage location details: the phenom 17 total length was measured to be ~160.3cm and the useable length was measured to be ~152.8cm which is within specification.No damages were found with the phenom hub.The phenom 17 catheter body was found to be kinked at ~8.3cm from distal tip.No damages or irregularities were found with the distal marker/tip.Testing/analysis (including sem reports): the phenom 17 catheter was flushed, water exited from the distal tip.The phenom-17 catheter was then tested with an in-house 0.016¿ mandrel.The mandrel successfully passed through the catheter hub, lumen and exited distal tip without issue.No other anomalies were observed.Conclusion: there was no indication that the event was related to a potential manufacturing issue.Based on the device analysis and reported information, the customer¿s report of ¿catheter resistance¿ was unable to be confirmed as no issue was found during catheter resistance testing with an in-house mandrel and the root cause could not be determined.Possible causes for ¿catheter resistance¿ are continuous flush rate too low, catheter or delivery system damage or insufficient delivery system hydration.It is possible the damage (kinked) found with the catheter body contributed to the resistance.Based on the device analysis and reported information, the customer¿s report of ¿catheter occlusion¿ could not be confirmed.No occlusion was found.H6.Coding updated based on analysis results.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATHERA
Type of Device
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key13636746
MDR Text Key286579489
Report Number2029214-2022-00331
Device Sequence Number1
Product Code KRA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K151638
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Model NumberFG11150-0615-2X
Device Catalogue NumberFG11150-0615-2X
Device Lot NumberOC20-022
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2022
Is the Reporter a Health Professional? No
Date Manufacturer Received03/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/05/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-