• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO UPPER G.I.SCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO UPPER G.I.SCOPE Back to Search Results
Model Number EG-2990I
Device Problems Fluid/Blood Leak (1250); Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2022
Event Type  malfunction  
Manufacturer Narrative
If additional information becomes available, a supplemental report will be filed with the new information.
 
Event Description
This event occurrence time is unknown.The date of occur is unkown.There was no report of patient harm.No knowledgement of adverse event refer to expiration date as complaint notification customer advised: resistance primary biopsy channel.
 
Manufacturer Narrative
Correction information g6: follow up #1 h2:if follow-up, what type? h3:device evaluated by manufacture.H6: coding changed based on the investigation result.Additional information h4:device manufacture date.Evaluation summary: we checked the returned unit and confirmed that the operation channel leak.Based on the result, we concluded that it was caused due to the excessive force applied on the operation channel.In addition, we confirmed that the light guide cable coating damage, and the remote control buttons disinfections damage; however, they are not the main cause, and/or irrelevant to the alleged complaint.The device has been repaired.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO UPPER G.I.SCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key13657037
MDR Text Key296675917
Report Number9610877-2022-00456
Device Sequence Number1
Product Code FDS
UDI-Device Identifier04961333129492
UDI-Public04961333129492
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131902
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEG-2990I
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/07/2022
Initial Date FDA Received03/02/2022
Supplement Dates Manufacturer Received02/07/2022
Supplement Dates FDA Received08/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-