Correction information g6: follow up #1 h2:if follow-up, what type? h3:device evaluated by manufacture.H6: coding changed based on the investigation result.Additional information h4:device manufacture date.Evaluation summary: we checked the returned unit and confirmed that the operation channel leak.Based on the result, we concluded that it was caused due to the excessive force applied on the operation channel.In addition, we confirmed that the light guide cable coating damage, and the remote control buttons disinfections damage; however, they are not the main cause, and/or irrelevant to the alleged complaint.The device has been repaired.
|