Catalog Number C58961
Device Problem False Negative Result (1225)
Patient Problem Insufficient Information (4580)
Event Date 11/11/2021
Event Type  malfunction  
Event Description
On (b)(6) 2022 the customer reported that on (b)(6) 2021, non-reactive covid igg (access sars-cov-2-igg, part number c58961 and lot number 125390) results were generated on the customer's access 2 (access 2 immunoassay analyzer, part number 81600n and serial number (b)(4)) for one patient.The non-reactive patient results were discordant to results generated using the diasorin sars-cov-2 trimeric s igg semi-quantitative result.There was no report of change to patient care or treatment in connection with this event.No hardware errors or issues with other assays were reported in conjunction with this event.System performance indicators such as system check, calibration and quality control were passing within specifications at the time of the event.No issues with sample integrity were reported by the customer.Sample collection, handling and processing information such as sample type, sample volume collected, sample quality, centrifugation time and speed, storage temperature and other information was not provided by the customer.
 
Manufacturer Narrative
The full identifier is (b)(6).The customer did not provide patient demographics such as date of birth, weight, ethnicity or race.The access sars-cov-2 igg assay was not returned for evaluation.There were no reports of system issues at the time of the event.There was no report of issues with other assays at the time of the event.No hardware errors or flags were reported in conjunction with the event.Sars-cov-2 is an enveloped non-segmented positive-sense rna virus.It has several structural proteins including spike (s), envelope (e), membrane (m) and nucleocapsid (n).The spike protein (s) contains a receptor binding domain (rbd) which is responsible for recognizing the cell surface receptor, angiotensin converting enzyme-2 (ace2).It is found that the rbd of the sars-cov-2 s protein strongly interacts with the human ace2 receptor leading to endocytosis into the host cells and viral replication.The access assay detects antibodies directed against the spike protein, which are more likely to neutralize the virus.No manufacturer guarantees both a specificity and sensitivity of 100%.The concentration of sars-cov-2 igg in a given specimen determined with assays from different manufacturers can vary due to differences in assay methods, diversity of antibodies and reagent specificity.Differences in each individual assay are expected.Different vaccines do not elicit the same immune response and each patient response is different therefore differences in patient responses are expected after vaccination.The access assay is not labeled for vaccine response detection.In conclusion, the cause of this event cannot be determined with the available information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCESS SARS-COV-2 IGG
Type of Device
IN VITRO DIAGNOSTIC DEVICES: CORONAVIRUS SEROLOGICAL REAGENTS
Manufacturer (Section D)
BECKMAN COULTER
1000 lake hazeltine drive
chaska MN 55318
Manufacturer (Section G)
BECKMAN COULTER
1000 lake hazeltine drive
chaska MN 55318
Manufacturer Contact
angela vettel
250 s. kraemer blvd.
brea, CA 92821
7149613625
MDR Report Key13660288
MDR Text Key289009478
Report Number2122870-2022-00012
Device Sequence Number1
Product Code QKO
UDI-Device Identifier15099590738648
UDI-Public(01)15099590738648(17)220731(11)211203(10)125390
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/03/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Catalogue NumberC58961
Device Lot Number125390
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/31/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age33 YR