Model Number 5450-35-500
Device Problem Loss of or Failure to Bond (1068)
Patient Problems Adhesion(s) (1695); Low Blood Pressure/ Hypotension (1914); Ischemia (1942); Pain (1994); Ambulation Difficulties (2544); Confusion/ Disorientation (2553); Swelling/ Edema (4577)
Event Date 03/17/2020
Event Type  Death  
Event Description
Patient received a right primary attune to treat severe end-stage arthritis.The patella was resurfaced and depuy cement x 2 was utilized.The surgeon notes that the patient had a preexisting left bka that was performed to treat a clotting issue.The procedure was completed without complications.Patient received a right knee revision to treat pain, swelling, and inability to walk secondary to aseptic loosening.The patient has extensive comorbidities.Because of this, he was given dialysis and extensive treatments for his preexisting clotting disorder before being approved for the revision and cleared for the procedure.Upon entering the joint, scar tissue was debrided from the joint.The tibial tray was grossly loose and debonded at the cement to implant interface and easily removed.The femoral component and patella were both loose at an unknown interface and revised.There was no reported product problem with the revised tibial insert.The patient was revised with competitor products.The procedure was completed without complications.Hours after the revision.The patient had an episode of hypotension and confusion and was transferred to the icu.Dialysis was attempted but was discontinued as the hypotension returned.Ct scan revealed significant ischemic and dead bowel.The family and physicians conversed, and corrective surgery was declined.Patient was given comfort measures and expired on the second post-operative day.There is insufficient information provided to determine the root cause of the death and the depuy synthes products cased or contributed to the reported death two days after the revision.Doi: (b)(6) 2016; dor: (b)(6) 2020; dod: (b)(6) 2020; right knee.
 
Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.Dmf#: (b)(4).Trade name gentamicin sulphate, active ingredient(s) gentamicin sulphate, dosage form - powder, strength 1.0g active in our cements.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (nc search) was performed for the finished device 545035500, and lot 8159147, no non-conformances / manufacturing irregularities related to the malfunction were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTSET GHV GENTAMICIN 40G
Type of Device
BONE CEMENT : BONE CEMENT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY CMW - 9610921
cornford rd
blackpool FY4 4 QQ
UK   FY4 4QQ
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13664733
MDR Text Key286553691
Report Number1818910-2022-04055
Device Sequence Number1
Product Code MBB
UDI-Device Identifier10603295174288
UDI-Public10603295174288
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K081163
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/03/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2017
Device Model Number5450-35-500
Device Catalogue Number545035500
Device Lot Number8159147
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received04/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/02/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ATTUNE MEDIAL DOME PAT 38MM.; ATTUNE PS FEM RT SZ 7 CEM.; ATTUNE PS RP INSRT SZ7 5MM.; ATTUNE RP TIB BASE SZ 6 CEM.; SMARTSET GHV GENTAMICIN 40G.
Patient Outcome(s) Death; Required Intervention;
Patient Age59 YR
Patient SexMale
Patient RaceBlack Or African American
-
-