APPLIED MEDICAL RESOURCES CB030, 5MMX35CM EPIX LAPA SCISSOR, 10/BX; LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES) Back to Search Results
Model Number CB030
Device Problems Dull, Blunt (2407); Failure to Cut (2587)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/11/2022
Event Type  malfunction  
Manufacturer Narrative
The event device is anticipated to be returned to applied medical for evaluation.A follow-up report will be provided upon completion of investigation.
 
Event Description
Procedure performed: inguinal hernia.Event description: scissors not sharp at the distal end which makes dissection during procedure more dangerous.This is a returning problem for dr [name] and was reported in an earlier cer.Patient ok! intervention: no information.Patient status: patient ok!.
 
Event Description
Procedure performed: inguinal hernia.Event description: scissors not sharp at the distal end which makes dissection during procedure more dangerous.This is a returning problem for dr [name] and was reported in an earlier cer.Patient ok! intervention: no information.Patient status: patient ok!.
 
Manufacturer Narrative
The event unit was returned to applied medical for evaluation.Testing was performed on the event unit, which confirmed that the returned unit was unable to cut consistently across the entire length of the blade.Engineering observed a small gap between the tips of the blades.Based on the condition of the event unit and description of the event, the scissors were unable to cut due to the gap that were observed between the tips of the blades on the returned unit.  however, the exact root cause of the gap could not be determined based on the evaluation of the returned unit.Applied medical continuously seeks to improve the form, function, and ease of use of its products.  as part of this process, applied medical is currently researching possible enhancements intended to further minimize the potential for this type of event to occur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CB030, 5MMX35CM EPIX LAPA SCISSOR, 10/BX
Type of Device
LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES)
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
aaron fulcher
22872 avenida empresa
rancho santa margarita, CA 92688
9497135765
MDR Report Key13691122
MDR Text Key291639610
Report Number2027111-2022-00447
Device Sequence Number1
Product Code HET
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K062169
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCB030
Device Catalogue Number101419701
Device Lot Number1428872
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/07/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-