• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO CUSTOM TUBING PACK; TUBING, PUMP, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO CUSTOM TUBING PACK; TUBING, PUMP, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HY10U10R4
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2022
Event Type  malfunction  
Event Description
Medtronic received information that prior to use of these custom tubing packs, the customer reported that the blood inlet connection site has been broken off of three fusion oxygenators.The custom tubing packs were replaced with new packs cases.There was no patient involvement so no adverse patient effects occurred.Additional information: the customer stated that they believe that the packaging design is the issue.There was no noticeable damage to shipping containers or to the packaging.Other devices in the same box/shipping container were not damaged.
 
Manufacturer Narrative
Product analysis: visual inspection showed that the qb-inlet port was broken off.The reason for return was confirmed.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Conclusion: the complaint is confirmed for a broken inlet port on the fusion oxygenator.Root cause investigation is still in progress.There were no patient/clinical safety issues reported.Trends for issues with this device are monitored.Correction b5: the event description has been updated to capture the fact that this issue occurred with one device rather than the three that were previously mentioned in the initial report.The broken inlet port was completely broken off prior to use.This would make the device unusable for the procedure, and so there is no risk of patient death or serious injury.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM TUBING PACK
Type of Device
TUBING, PUMP, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key13706127
MDR Text Key290275318
Report Number9612164-2022-00920
Device Sequence Number1
Product Code DWE
UDI-Device Identifier00763000542139
UDI-Public00763000542139
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K171308
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/18/2023
Device Model NumberHY10U10R4
Device Catalogue NumberHY10U10R4
Device Lot Number222374724
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2022
Is the Reporter a Health Professional? No
Date Manufacturer Received03/30/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/07/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-