• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL ENSITE VELOCITY¿ SYSTEM VELOCITY AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ENSITE VELOCITY¿ SYSTEM VELOCITY AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number 100014514
Device Problems Sparking (2595); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/24/2022
Event Type  malfunction  
Event Description
During a procedure, when removing the power cord from the amplifier, a spark was noted.Upon removal, it was noted that one of the prongs from the amplifier had detached and was stuck in one of the connecting slots of the power cord.The procedure was cancelled as a result of the issue.There were no adverse consequences to any users or the patient.
 
Manufacturer Narrative
Additional information: d9, g3, h2, h3.One ensite velocity¿ amplifier was received for evaluation.Visual inspection revealed the front panel ports, chassis, and labels were free of physical damage.Visual inspection on the rear of the amplifier shows that one of the incoming alternating current posts was missing from the power input module (pn: 100005066), which duplicates the reported symptom.Using a known good power input module, the amplifier was able to be evaluated further.Power was applied to the amplifier and the amplifier passed power-on-self-test (post).The amplifier went status ready ¿green¿ and communicated with the test ensite computer.Evaluation of the board status showed all boards status are green indicating passing results.Evaluation of the logs revealed showed no reportable symptoms.The reported event was able to be duplicated by visual inspection.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information provided to abbott, images returned, and the investigation performed, the root cause was attributed to physical damage to the power input module.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE VELOCITY¿ SYSTEM VELOCITY AMPLIFIER
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key13725236
MDR Text Key287194643
Report Number2184149-2022-00071
Device Sequence Number1
Product Code DQK
UDI-Device Identifier05414734210713
UDI-Public(01)05414734210713(10)5263140
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number100014514
Device Catalogue Number100014514
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/11/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-