Model Number CI-1500-04 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Erythema (1840); Hypersensitivity/Allergic reaction (1907); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 03/04/2022 |
Event Type
Injury
|
Event Description
|
The recipient reportedly presented with redness and swelling at the skin flap.The reaction is believed to be an allergic response to the equipment.The recipient's device was explanted.The recipient will be reimplanted at a later date.
|
|
Manufacturer Narrative
|
The recipient has reportedly healed following surgery.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
A review of the device history record was completed upon request by the nca and no anomalies were noted.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed tool damage to the top cover, as well as a severed electrode.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical tests performed.The device passed the mechanical test performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|