U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION 213, LLC 0113 CHLORAPREP ONE STEP HI-LITE ORANGE; 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION 213, LLC 0113 CHLORAPREP ONE STEP HI-LITE ORANGE; 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL Back to Search Results
Catalog Number 930815
Device Problem Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/26/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(6) initial emdr submission.A follow up emdr will be submitted if additional information becomes available.(b)(4).
 
Event Description
It was reported that chloraprep leaking in pack.Verbatim: received via: email.External reference # (b)(4).# of occurrences -1.
 
Event Description
It was reported that chloraprep leaking in pack verbatim: received via : email.External reference # (b)(4).# of occurrences -1.Reported issue: chloraprep leaking in pack.
 
Manufacturer Narrative
Photos were available for evaluation.Visual examination of the photos show the foam tip and lidding tinted orange in a kit which verifies the failure mode of pre-activation.When pressure is applied to the wings by a pinching force with the fingers, this activates the applicator, breaking the glass ampoule and releasing the chloraprep solution onto the foam tip.The most probable root cause can be attributed to post manufacturing handling, which can provide enough force/impact to activate and break the glass ampoule.Due to the nature of glass, it is possible to have an activated applicator and/or broken ampoule if the applicator undergoes excessive handing.A production record review was completed for batch/lot 1191738 and no non conformances were identified during the manufacturing of this lot.No further actions are required at this time.This failure will continue to be tracked and trended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHLORAPREP ONE STEP HI-LITE ORANGE
Type of Device
2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL
Manufacturer (Section D)
CAREFUSION 213, LLC 0113
1550 northwestern dr
el paso TX 79912
Manufacturer (Section G)
CAREFUSION, INC
75 n. fairview drive
vernon hills IL 60061
Manufacturer Contact
anna wehrheim
75 n. fairview drive
vernon hills, IL 60061
8015652341
MDR Report Key13757777
MDR Text Key287115821
Report Number3004932373-2022-00071
Device Sequence Number1
Product Code KXG
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number930815
Device Lot Number1191738
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/25/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-