Model Number 893137
Device Problem Inaccurate Flow Rate (1249)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/17/2022
Event Type  malfunction  
Manufacturer Narrative
During laboratory analysis, the product surveillance technician (pst) installed the flowmeter into a lab use only (luo) electronic patient gas system (epgs).Verification/release testing was conducted and the pst observed the flowmeter to function as intended and pass all testing.
 
Event Description
The service repair technician (srt) reported that during testing of the device at the service center, the flowmeter failed accuracy testing.There was no patient involvement.
 
Manufacturer Narrative
Updated block: h6 the reported complaint was confirmed.Per supplier evaluation, the device was out of tolerance and found to have values 6% high.The device was shipped with system gain of 42.88 and recalibrated to 40.21.The higher system gain was evidence there was a leak during calibration and the direct material root cause of the leak during the calibration was determined to be a fitting used in the test stand between the reference and the device that was missing an o-ring.The higher system gain was evidence that the calibration process compensated for the leak present during the device's original calibration.The recalibration with lower system gain with no leak present further verified this.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key13760033
MDR Text Key287284243
Report Number1828100-2022-00107
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number893137
Device Catalogue Number893137
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/13/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-