MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 397002-001
Device Problems Battery Problem (2885); Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2022
Event Type  malfunction  
Manufacturer Narrative
The companion 2 driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4) initial.
 
Event Description
The companion 2 driver was not supporting a patient.The supplier, a syncardia authorized warehouse, reported that the emergency battery of the companion 2 driver did not charge.
 
Manufacturer Narrative
The customer-reported issue was confirmed upon review of the patient file which showed the driver had two emergency battery error alarms.The issue was replicated during incoming functional testing, in which multiple steps did not pass due to the emergency battery being unable to power driver and displaying an emergency battery error alarm.The root cause of the companion 2 driver's emergency battery not charging was determined to be a malfunction of the emergency battery.The battery was found to have a permanent fault status which prohibits the battery from charging or discharging.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its investigation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
kerri hensley
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key13763220
MDR Text Key287183612
Report Number3003761017-2022-00036
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Administrator/Supervisor
Type of Report Followup
Report Date 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received03/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/07/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-