Model Number 397002-001 |
Device Problems
Battery Problem (2885); Power Problem (3010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/11/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The companion 2 driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4) initial.
|
|
Event Description
|
The companion 2 driver was not supporting a patient.The supplier, a syncardia authorized warehouse, reported that the emergency battery of the companion 2 driver did not charge.
|
|
Manufacturer Narrative
|
The customer-reported issue was confirmed upon review of the patient file which showed the driver had two emergency battery error alarms.The issue was replicated during incoming functional testing, in which multiple steps did not pass due to the emergency battery being unable to power driver and displaying an emergency battery error alarm.The root cause of the companion 2 driver's emergency battery not charging was determined to be a malfunction of the emergency battery.The battery was found to have a permanent fault status which prohibits the battery from charging or discharging.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its investigation and is closing this file.(b)(4) follow-up report 1.
|
|