• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION BIPAP AUTOSV; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION BIPAP AUTOSV; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX900H11C
Device Problem Degraded (1153)
Patient Problems Dyspnea (1816); Sore Throat (2396); Nasal Obstruction (2466)
Event Date 02/21/2022
Event Type  malfunction  
Event Description
The manufacturer received information alleging an issue related to a bipap device's sound abatement foam.The patient has alleged to experiencing nasal and throat soreness.There was no report of patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
 
Manufacturer Narrative
The manufacturer previously reported an issue with the device's sound abatement foam.The manufacturer received additional information alleging difficulty breathing/short of breath.There was no allegation of serious or permanent harm or injury.
 
Manufacturer Narrative
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging nasal and throat soreness, difficulty breathing/short of breath related to a bipap device's sound abatement foam.There was no report of patient harm or injury.Repeated attempts to have the device and components returned for evaluation and investigation were unsuccessful.The manufacturer believes they will be unable to gather additional information.The manufacturer is submitting an updated report at this time.If pertinent information becomes available to the manufacturer at a later date, a follow-up report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION BIPAP AUTOSV
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh
pittsburgh, PA 15206
2673970028
MDR Report Key13790457
MDR Text Key297229970
Report Number2518422-2022-10633
Device Sequence Number1
Product Code MNS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090539
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 07/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX900H11C
Device Catalogue NumberDSX900H11C
Was Device Available for Evaluation? No
Date Manufacturer Received07/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/21/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberRES 88058
Patient Sequence Number1
-
-