Model Number PL569T
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/12/2022
Event Type  Injury  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Event Description
It was reported that there was an issue with pl569t - ligating clips m/l 12/box.According to the complaint description, the disposable material fell apart while being used in the operating room during the operation in the abdomen.It has happened once before recently, but the lot number was not noted.There is no sample available and there was no patient damage.Possible damage could be that the individual parts may remain in the patient's abdomen, and cause inflammation.The sales representative was at the customer's site.At first, she thought the safety ring might be missing.But it wasn't the stock that fell out, it was the magazine.It also disintegrated into its individual parts.The doctor had to fish out clips.An additional medical intervention was necessary.Additional information was not provided nor available.Additional patient information is not available.The adverse event is filed under aesculap (b)(4) reference (b)(4).
 
Manufacturer Narrative
Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 3(5) x probability of occurrence 1(5)) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigations results a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGATING CLIPS M/L 12/BOX
Type of Device
LIGATION / VESSEL CLIPS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key13791148
MDR Text Key288620783
Report Number9610612-2022-00076
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K962493
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/17/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPL569T
Device Catalogue NumberPL569T
Device Lot Number52735023
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/18/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-