• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER; ELECTRODE, PACEMAKER, TEMPORARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER; ELECTRODE, PACEMAKER, TEMPORARY Back to Search Results
Model Number D97120F5
Device Problem Pacing Inadequately (1442)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/24/2022
Event Type  malfunction  
Manufacturer Narrative
It was further informed that the device was not available for evaluation.Without return of the product, edwards is unable to perform a complete investigation of the reported event.It is not possible to determine what factors may have contributed to it, and therefore no actions could be planned.Based on further engineering investigation, as part of the manufacturing process the units go through an electrode and electrical inspection process.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
As reported, during a valve-in-valve intervention, this swan-ganz bipolar pacing catheter connected to an external osypka pacer did not pace.The monitor was changed but the issue remained.Once the swan ganz catheter was changed the issue was solved.There was no allegation of patient injury.The device was available for evaluation.
 
Manufacturer Narrative
The product is expected to be returned for analysis; however, it has not yet been received.Upon the return of the product a supplemental report will be sent with the investigation results.The lot number for this device was not supplied; therefore, further review of the related manufacturing records could not be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ BIPOLAR PACING CATHETER
Type of Device
ELECTRODE, PACEMAKER, TEMPORARY
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
samantha eveleigh
1 edwards way
irvine, CA 92614
9492503939
MDR Report Key13800650
MDR Text Key290464155
Report Number2015691-2022-04473
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K813521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/17/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberD97120F5
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexFemale
-
-