ABBOTT GMBH MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM Back to Search Results
Model Number 03P6822
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/25/2022
Event Type  malfunction  
Event Description
The customer reported a (b)(6) magnesium result generated on the architect c4000 analyzer on a patient.Results provided: (b)(6) 2022 (b)(6).Normal range: 1.6-2.6 mg/dl.No impact to patient management was reported.
 
Manufacturer Narrative
Ticket searches and performance testing could not be performed as the likely cause lot number is unknown.There are no trends for the product related to patient results.No customer returns were available for evaluation.A review of overall customer field data was performed to assess the performance of the architect magnesium assay and is within the established baselines confirming no systemic issues in the field.A review of the manufacturing documentation did not identify any issues associated with the customers observation.A review of labeling concluded that the issue is sufficiently addressed.No product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGNESIUM
Type of Device
PHOTOMETRIC METHOD, MAGNESIUM
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key13801480
MDR Text Key289484985
Report Number3002809144-2022-00097
Device Sequence Number1
Product Code JGJ
UDI-Device Identifier00380740161620
UDI-Public00380740161620
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181748
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/17/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number03P6822
Device Catalogue Number03P68-22
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ARC C4000 INTGR, 02P24-40, C401897
-
-