Model Number GIF-H260Z |
Device Problems
Contamination (1120); Device Damaged by Another Device (2915)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/15/2022 |
Event Type
malfunction
|
Event Description
|
As reported for this event by the customer, during a gastroscopy procedure, a syringe needle was stuck at the channel tube of the device.The procedure was completed with the same device with no more than 15 minutes delay.There is no reported harm or adverse impact to the patient.
|
|
Manufacturer Narrative
|
Full name of the facility is (b)(6) hospital of traditional (b)(6) medicine.This device is not sold in the us but a similar device is.The device is returned but the device evaluation is not yet completed.As such a definitive root cause of the reported complaint cannot be determined at this time.Supplemental report(s) will be filed as any relevant new information becomes available.
|
|
Manufacturer Narrative
|
This supplemental report is to inform that upon further review, this is not a reportable malfunction.Per the legal manufacturer, there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.
|
|
Event Description
|
Addendum (b)(6) 2022: the procedure was diagnostic.The needle got stuck in the device when the device was sending water during the procedure.Details of how it was stuck are not provided.The needle appeared at the mouth of the channel and could not be removed after inspection.As such, the device was sent in for repair.The device was inspected prior to use; inspection results were a normal image and no issues with the air/water.
|
|