Model Number GIF-H260Z
Device Problems Contamination (1120); Device Damaged by Another Device (2915)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/15/2022
Event Type  malfunction  
Event Description
As reported for this event by the customer, during a gastroscopy procedure, a syringe needle was stuck at the channel tube of the device.The procedure was completed with the same device with no more than 15 minutes delay.There is no reported harm or adverse impact to the patient.
 
Manufacturer Narrative
Full name of the facility is (b)(6) hospital of traditional (b)(6) medicine.This device is not sold in the us but a similar device is.The device is returned but the device evaluation is not yet completed.As such a definitive root cause of the reported complaint cannot be determined at this time.Supplemental report(s) will be filed as any relevant new information becomes available.
 
Manufacturer Narrative
This supplemental report is to inform that upon further review, this is not a reportable malfunction.Per the legal manufacturer, there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.
 
Event Description
Addendum (b)(6) 2022: the procedure was diagnostic.The needle got stuck in the device when the device was sending water during the procedure.Details of how it was stuck are not provided.The needle appeared at the mouth of the channel and could not be removed after inspection.As such, the device was sent in for repair.The device was inspected prior to use; inspection results were a normal image and no issues with the air/water.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13801736
MDR Text Key287614867
Report Number8010047-2022-04595
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 04/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/17/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-H260Z
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/01/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CV-290.
-
-