The device was returned for analysis to boston scientific.Per visual inspection the device presents a bent/kink in the proximal part of the tip.Also, there is a cut in the distal part of the catheter shaft.Functional test showed that the steering knob and the tension control knob functioned properly on both lock and unlock positions.No abnormal resistance was felt when actuating the steering mechanism.There is an open in the tip and second ring after continuity test finding.Per dissection test the cut in the catheter shaft was confirmed affecting the tip and the ring 2.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.
|
Reportable based on device analysis completed on 24feb2022.It was reported that during a cardiac radiofrequency ablation procedure, a blazer ii catheter had poor contact and could not discharge.There were no issues regarding temperature.The catheter was discharged intermittently and could not be used normally.No error messages.Temperature control mode was in use.Procedure was completed with another of same device.No patient injury reported.Device was returned to boston scientific for analysis.However, device analysis revealed that there is a cut in the distal part of the catheter shaft.
|