In addition to the power inlet and internal connectors being loose, service confirmed that the power switch was not working.Service also found that the suction feet had lost suction, the front connector was loose, the connector air pipe rubber was worn, and the switch cap was damaged.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The dhr was unable to be reviewed for this device since the device is over 15 years old.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, a definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.
|