• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG COVERA VASCULAR COVERED STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG COVERA VASCULAR COVERED STENT Back to Search Results
Model Number AVSM09040
Device Problems Obstruction of Flow (2423); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Stenosis (2263)
Event Date 02/08/2022
Event Type  Injury  
Event Description
It was reported through the results of a clinical trial, that approximately nine months post index procedure, stenosis in target lesion was observed and standard pta was used to successfully treat the target lesion.The current status of the subject is not provided.
 
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Expiry date: 08/2022.
 
Manufacturer Narrative
Manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was not returned to the manufacturer for evaluation and images were not provided for review.Based on the provided information, the investigation is inconclusive for reported issue.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.Under potential complications the instructions for use state 'new lesions in the access circuit requiring reintervention, thrombotic occlusion, restenosis of the target lesion requiring reintervention, pseudoaneurysm, (¿)'.In regards to pta the instructions for use state: 'pre-dilate the stenosis with a pta balloon catheter of appropriate length and diameter for the lesion to be treated'.And 'post dilate the covered stent with an angioplasty balloon sized appropriately as to ensure complete wall apposition to the reference vessel.Avoid balloon dilation in the healthy, non-stenosed segment of the vein'.Holding and handling of system was found described, in particular the instructions for use state: 'do not touch the distal catheter assembly (i.E.The dark brown catheter segment) during covered stent deployment since this may interfere with covered stent deployment and may lead to misplacement'.:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Device not returned.
 
Event Description
It was reported through the results of a clinical trial, that approximately nine months post index procedure, stenosis in target lesion was observed and standard pta was used to successfully treat the target lesion.The current status of the patient is not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COVERA VASCULAR COVERED STENT
Type of Device
VASCULAR COVERED STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13837783
MDR Text Key287578136
Report Number9681442-2022-00081
Device Sequence Number1
Product Code PFV
UDI-Device Identifier00801741106330
UDI-Public(01)00801741106330
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P170042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Study
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAVSM09040
Device Catalogue NumberAVSM09040
Device Lot NumberANEV0525
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age67 YR
Patient SexFemale
Patient Weight53 KG
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-