Patient on medical stepdown unit had rapid response called for hypotension and oxygen desaturations.Decision made to transfer patient to icu.As staff were preparing to transfer the patient, they switched the patient from the wall oxygen supply to an liv (linde integrated valve) portable oxygen tank.Staff turned the top valve on, heard a release of air, and thought oxygen was flowing.Within a minute or so, the patient's oxygen saturations dropped to the 50s.The patient went into cardiorespiratory arrest, cpr was started, and code blue was activated.Once rosc (return of spontaneous circulation) achieved, the rn checked o2 tank and found the second valve (the one that turns the oxygen on and off) which is located on the side of the tank was in the "off" (red) position.The linde liv integrated valve oxygen system (portable oxygen tank with regulator) has 2 valves.The regulator valve is on top and on/off valve is on the side.Both valves need to be open for oxygen to flow.When you open the top valve, there is a hissing noise which can mislead the user into thinking that the oxygen is flowing and fully turned on.Subsequently, they may not open the side valve because they think the oxygen is already "on." we have had a number of near misses since this tank was introduced to our hospital in (b)(6) 2017.We believe the design of this tank presents a serious safety concern.
|