Model Number A-TCSE-D
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2021
Event Type  malfunction  
Manufacturer Narrative
One uni-directional, curve d, tacticath sensor enabled contact force ablation catheter was received for evaluation.Microscopic analysis of the distal tip revealed no deformity or damage to electrode ring 2.Further investigation revealed electrode ring 2 was displaced distally.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the displaced electrode ring 2 remains unknown.
 
Event Description
This report is to advise of an event noted during analysis which revealed a displaced electrode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ UNID CURVE D
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key13891553
MDR Text Key287911060
Report Number3008452825-2022-00129
Device Sequence Number1
Product Code OAE
UDI-Device Identifier05415067027689
UDI-Public05415067027689
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2023
Device Model NumberA-TCSE-D
Device Catalogue NumberA-TCSE-D
Device Lot Number8125074
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/06/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-