OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number CON-HL-90
Device Problem Electrical Power Problem (2925)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/25/2022
Event Type  malfunction  
Manufacturer Narrative
One device was received for evaluation in relation to the reported event.Evaluation of the device is currently in progress.Once the evaluation is complete, a follow-up report with evaluation results will be submitted.
 
Event Description
It was reported that the device was drawing over two amps in current.There was no additional information provided.
 
Manufacturer Narrative
No patient involvement.
 
Event Description
It was reported that the device was drawing over two amps in current.Event date updated.No patient involvement.
 
Manufacturer Narrative
Other text: h6: event problem and evaluation codes: updated.H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection found the device with corroded drain fitting, cracked tank cover, broken pole clamp, outdated pcb (printed circuit board) and power switch.The customer stated problem was duplicated.No action taken due to the age and condition of the device.It is deemed beyond economical repair and will be scrapped.The cause of the reported problem could not be determined.A manufacturing dhr review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
receiving dock
minneapolis, MN 55442
MDR Report Key13892153
MDR Text Key287899021
Report Number3012307300-2022-05255
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/24/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/28/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-