Catalog Number CON-HL-90 |
Device Problem
Electrical Power Problem (2925)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/25/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
One device was received for evaluation in relation to the reported event.Evaluation of the device is currently in progress.Once the evaluation is complete, a follow-up report with evaluation results will be submitted.
|
|
Event Description
|
It was reported that the device was drawing over two amps in current.There was no additional information provided.
|
|
Manufacturer Narrative
|
No patient involvement.
|
|
Event Description
|
It was reported that the device was drawing over two amps in current.Event date updated.No patient involvement.
|
|
Manufacturer Narrative
|
Other text: h6: event problem and evaluation codes: updated.H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection found the device with corroded drain fitting, cracked tank cover, broken pole clamp, outdated pcb (printed circuit board) and power switch.The customer stated problem was duplicated.No action taken due to the age and condition of the device.It is deemed beyond economical repair and will be scrapped.The cause of the reported problem could not be determined.A manufacturing dhr review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.
|
|