Upon receipt at our post market quality assurance laboratory, visual inspection of the lead found that the helix mechanism was extended and dried blood was noted in the helix housing.A stylet passed through the lead, but with resistance felt due to blood/body fluid which broke apart as the stylet was pushed forward.Based upon the clinical observations and the laboratory findings, we believe that body fluid inside the helix housing may have contributed to the irregularity in helix function and stylet insertion.Then testing was completed to assess lead electrical performance and inner/outer insulation integrity.Measurements throughout these tests were within normal limits.Laboratory analysis did not identify any lead characteristics that would have caused or contributed to the reported clinical observations regarding loss of capture allegation.
|
It was reported that during implant procedure this right ventricular (rv) lead was an attempt due to dislodgment, placement difficulty and loss of capture with asystole less than 2 seconds.It was noted that the lead lost slack and got dislodged after surgery.Subsequently, the physician tried to repositioned the lead on the same day, nonetheless the stylet was difficult to insert and the helix did not extend after stylet insertion.This rv lead successfully replaced.No additional adverse patient effects were reported.
|