|
Model Number N/A |
Device Problems
Loose or Intermittent Connection (1371); Excessive Heating (4030)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/09/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
|
Event Description
|
It was reported this evac was not connecting to the carts.The nose board on the coupler was burnt.No harm, injury, or adverse event to the operator.The timing was during cleaning.It was reported that no further information is available.
|
|
Manufacturer Narrative
|
This follow-up is being submitted to relay additional information.Updated: b4, b5, d4, g3, g6, h1, h2, h3, h4,h6, and h10.The results of the investigation are as follows: review of the most recent repair record determined the nose board was damaged due to water and was replaced.-the event is confirmed.-the device was tested by third party and found to be functioning to specifications prior to release to the customer.-a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that no further information is available.
|
|
Search Alerts/Recalls
|
|
|