Model Number N/A
Device Problems Loose or Intermittent Connection (1371); Excessive Heating (4030)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
It was reported this evac was not connecting to the carts.The nose board on the coupler was burnt.No harm, injury, or adverse event to the operator.The timing was during cleaning.It was reported that no further information is available.
 
Manufacturer Narrative
This follow-up is being submitted to relay additional information.Updated: b4, b5, d4, g3, g6, h1, h2, h3, h4,h6, and h10.The results of the investigation are as follows: review of the most recent repair record determined the nose board was damaged due to water and was replaced.-the event is confirmed.-the device was tested by third party and found to be functioning to specifications prior to release to the customer.-a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that no further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVAC STATION
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer (Section G)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key13925213
MDR Text Key288712379
Report Number0001954182-2022-00017
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00889024466012
UDI-Public(01)00889024466012
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172481
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00514010900
Device Lot Number0024772
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/20/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-