Catalog Number 5C4482
Device Problem Material Separation (1562)
Patient Problem Peritonitis (2252)
Event Date 03/01/2022
Event Type  Injury  
Event Description
A peritoneal dialysis (pd) patient experienced a separation between the female connector and the main body of a minicap transfer set while "disconnecting".The separation was further specified as ¿the entire dark blue piece unscrews from the light blue piece when patient disconnects from tubing¿.Subsequently the patient developed peritonitis.The cause of the separation was not reported, and no leak was noted.It was not reported if the patient was hospitalized for the event.The patient was treated with unspecified antibiotics for the event.The nurse noted after the first transfer set change, the device was ¿defective¿, so a second transfer set change was performed.At the time of this report, the patient outcome and action with pd therapy was not reported.No additional information is available.
 
Manufacturer Narrative
The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
The actual device was not available; however, a photograph of the sample was provided for evaluation.The visual inspection by the naked eye of the provided picture was performed and showed the separation between female connector and main body.The reported condition was verified.The cause of the condition was determined to be a manufacturing issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP TRANSFER SET
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13932442
MDR Text Key288100046
Report Number1416980-2022-01368
Device Sequence Number1
Product Code KDJ
UDI-Device Identifier00085412007731
UDI-Public(01)00085412007731
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152675
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number5C4482
Device Lot NumberH21J12075
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/12/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-