Model Number 172015 |
Device Problems
No Device Output (1435); Retraction Problem (1536)
|
Patient Problems
Hemorrhage/Bleeding (1888); Unspecified Tissue Injury (4559)
|
Event Date 03/04/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a laparoscopic procedure, at the beginning, the safety device did not work when the device was inserted; the needle did not retract and was exposed inside the abdominal cavity, damaging the patient's liver.A new device was used to resolve the issue.
|
|
Event Description
|
According to the reporter, during a laparoscopic cholecystectomy, at the beginning, the safety device did not work when the device was inserted; the needle did not retract and was exposed inside the abdominal cavity, damaging the patient's liver.There was a limited liver bleeding.To resolve that, the surgeon coagulated the tissue with no consequences.A new device was used to resolve the issue.
|
|
Manufacturer Narrative
|
Additional info: b5, g3, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a laparoscopic cholecystectomy, at the beginning, the safety device did not work when the device was inserted; the needle did not retract and was exposed inside the abdominal cavity, damaging the patient's liver.It was noted that the protector spring lost sensitivity as if it had not triggered, there was no "click" when it was entered.There was a limited liver bleeding.To resolve that, the surgeon coagulated the tissue with no consequences.A new device was used to resolve the issue.
|
|
Manufacturer Narrative
|
Correction: b5, g3, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: b2, g1 (manufacturer name, mfr contact first name, last name, street 1, mfr city, region, postal code, email, phone number), g3 correction: b5 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a laparoscopic cholecystectomy, at the beginning, the safety device did not work when the device was inserted; the needle did not retract and was exposed inside the abdominal cavity, damaging the patient's liver.It was noted that the protector spring lost sensitivity as if it had not triggered, and there was no "click" when it was entered.There was a limited liver bleeding.The issue happened with two devices.To resolve that, the surgeon coagulated the tissue with no consequences.A new device was used to resolve the issue.
|
|
Search Alerts/Recalls
|
|