Qn#(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.A device history record review was performed, and a potentially relevant finding was identified.A non-conformance was initiated for batches 13c21h1827, 13c21h1829, and 13c21b1929, to address the issue of a leaking extension line.Despite this, based on the customer report for this complaint that they "lost control of the wire", it was concluded that this is not the same issue being addressed within the relevant finding.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|