• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDIVATORS, INC. ADVANTAGE PLUS ENDOSCOPE REPROCESSING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVATORS, INC. ADVANTAGE PLUS ENDOSCOPE REPROCESSING SYSTEM Back to Search Results
Model Number 1-2-510.003
Device Problem Smoking (1585)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/02/2022
Event Type  malfunction  
Manufacturer Narrative
Medivators has requested to perform an inspection of the unit.The user facility stated the unit is not available for inspection as it has been removed from service while the facility plumbing is being repaired.The unit was installed in 2014 and is not under medivators service agreement for maintenance activities; the user facility is responsible for all maintenance activities.A follow-up mdr will be submitted should additional information become available.
 
Event Description
The user facility reported that smoke was emitting from the back of their advantage plus endoscope reprocessing system.No report of injury.
 
Manufacturer Narrative
Following completion of the repairs to the facility plumbing, the user facility contacted medivators to perform an inspection of the advantage plus endoscope reprocessing system.A medivators service technician arrived onsite to inspect the unit and found a connection on the hard drive was burnt resulting in the reported smoke.The technician observed that the vent holes on the back of the pc were blocked.Without proper ventilation, the hard drive can begin to overheat which likely caused to damage to the hard drive connection.The advantage plus endoscope reprocessing system states (13), "do not block ventilation openings." the technician replaced the hard drive and connections, tested the unit, confirmed it to be operating according to specification, and returned it to service.The unit was installed in 2014 and is not under medivators service agreement for maintenance activities; the user facility is responsible for all maintenance activities.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTAGE PLUS ENDOSCOPE REPROCESSING SYSTEM
Type of Device
ENDOSCOPE REPROCESSING SYSTEM
Manufacturer (Section D)
MEDIVATORS, INC.
14605 28th avenue n
plymouth MN 55447
Manufacturer (Section G)
MEDIVATORS, INC.
14605 28th avenue n
plymouth MN 55447
Manufacturer Contact
daniel davy
9800 59th avenue n
plymouth, MN 55442
4403927453
MDR Report Key13971030
MDR Text Key298136543
Report Number2150060-2022-00021
Device Sequence Number1
Product Code FEB
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/31/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1-2-510.003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/14/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-