• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. UNKN COBLATOR II SURGERY SYSTEM AND WAND; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. UNKN COBLATOR II SURGERY SYSTEM AND WAND; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number UNKN04300102
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Abrasion (1689); Bruise/Contusion (1754)
Event Date 01/12/2022
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).Article citation: zhang hl, ding hf, sun w, huang zl, hu y.[clinical effect of nasal endoscope combined with supporting laryngoscope surgery in the treatment of polyps of vocal cord and its influence on voice function of patients].Lin chung er bi yan hou tou jing wai ke za zhi.2019 mar;33(3):272-274.Chinese.Doi: 10.13201/j.Issn.1001-1781.2019.03.021.Pmid: 30813701.
 
Event Description
It was reported that on literature review "clinical study of nasal endoscopy and microscopy of vocal cord masses with support laryngoscope", 13 patients had a postoperative soft palate abrasion and submucosal bruising of the posterior pharyngeal wall after procedure using a coblator ii surgery system.Unknown how the events were treated.Patients outcome are unknown.No further information is available.
 
Manufacturer Narrative
Internal complaint reference (b)(4).H10 h3, h6 the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.Insufficient product identification information was provided and thus a manufacturing record, complaint history, device labeling/instructions for use, risk management and capa, nc, pra/hhe review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.Per subsequent e-mail, no further information is available.Without clinically relevant patient-specific supporting documentation, a thorough medical investigation could not be performed.The root cause and/or patient outcome beyond that which was documented in the article could not be confirmed nor concluded; therefore, no further medical assessment is warranted at this time.Should any additional relevant clinical information be provided, this case would be re-assessed.There was no relationship found between the device and the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN COBLATOR II SURGERY SYSTEM AND WAND
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13995594
MDR Text Key288595771
Report Number3006524618-2022-00165
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K202006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Study,Literature,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/04/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNKN04300102
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-