U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO ENTERRA; INTESTINAL STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 3116
Device Problems Break (1069); Degraded (1153); High impedance (1291)
Patient Problems Calcium Deposits/Calcification (1758); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/31/2022
Event Type  Injury  
Manufacturer Narrative
Product id 435135 lot# serial# (b)(4) implanted: (b)(6) 2007 explanted: (b)(6) 2022 product type: lead.Product id 435135 lot# serial# (b)(4) implanted: (b)(6) 2007 explanted: (b)(6) 2022 product type: lead.Information references the main component of the system.Other relevant device(s) are: product id: 435135, serial/lot #: (b)(4) ubd: 23-may-2008, udi#: (b)(4).Product id: 435135, serial/lot #: (b)(4), ubd: 30-may-2008, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was implan ted with an implantable neurostimulator (ins) for gastric stimulation.It was reported that while doing a battery change for battery depletion, hcp completed interrogating the device and found the impedance out of range (greater than 800).She decided to replace the leads as well at that time as the leads were 15 years old.She did so and the impedance returned to normal.The issue was resolved at the time of the report.The device was discarded.
 
Event Description
Additional information was received from a hcp via a manufacturer representative (rep).The rep reported that the per the doctor, "the device was very old- implanted at another hospital in 2007 i believe.I¿m sure that there were microfractures.There was corrosion around the battery and calcification around this and the leads.".
 
Manufacturer Narrative
Continuation of d10: product id 435135 serial# (b)(6) implanted: (b)(6) 2007 explanted: (b)(6) 2022 product type lead product id 435135 serial# (b)(6) implanted: (b)(6) 2007 explanted: (b)(6) 2022 product type lead medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos,pr MN 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos,pr MN 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key14015172
MDR Text Key290540598
Report Number2182207-2022-00579
Device Sequence Number1
Product Code LNQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/14/2007
Device Model Number3116
Device Catalogue Number3116
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/05/2022
Date Device Manufactured05/23/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age36 YR
Patient SexFemale
Patient Weight59 KG
-
-