Model Number 8002950 |
Device Problem
No Flow (2991)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that there was not any fluid flowing.No additional information.
|
|
Manufacturer Narrative
|
Patient identifier removed as no patient was involved.
|
|
Event Description
|
Additional information was received including the event date, and that no patient was involved as the issue was discovered during quarterly preventative maintenance.
|
|
Manufacturer Narrative
|
Manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.Visual inspection was performed.Visual inspection found that getting a disposable alarm that cannot be resolved by reseating the test filter.This is a condition that does not allow fluid to circulate.On closer inspection, the air detector sensor appears to be damaged.The root cause of the reported issue was found to be the damaged air detector sensor.This issue will be monitored for any increase in occurrence.If the occurrence of this issue increases significantly, additional corrective actions will be taken.
|
|