MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 895208
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2022
Event Type  malfunction  
Event Description
During testing of the device at the service center, the service repair technician (srt) reported that one of the elbow fittings from the electronic patient gas system (epgs) oxygen (o2) sensor upgrade kit was leaking air during the leak test above 10 cubic centimeters per minute (cc/min).There was no patient involvement.
 
Manufacturer Narrative
Per the srt, in the process of troubleshooting the thread was found to be distorted on the machined housing at the fitting.
 
Manufacturer Narrative
Updated block: h6 the reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key14017414
MDR Text Key288674015
Report Number1828100-2022-00139
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number895208
Device Catalogue Number895208
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/09/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-