SMITH & NEPHEW, INC. JRN II BCS LCK FEM IMPLANT IMPACT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
Model Number 74012812 |
Device Problems
Crack (1135); Scratched Material (3020)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/15/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference: case(b)(4).
|
|
Event Description
|
It was reported that, the plastic tips of a jrn ii bcs lck fem implant impact are cracked and scratched.It is unknown if this was noticed during surgery or not, therefore, patient involvement has not been confirmed.
|
|
Manufacturer Narrative
|
H3, h6: the associated device was returned and evaluated.A visual inspection was conducted and confirmed the impactor cam arms are cracked and have multiple scratches.The visual also confirms cement is found to be between the cam arm and the metal part of the device.This device exhibits signs of significant wear/ usage.For cleaning procedures please refer to smith and nephew¿s recommended cleaning methods given in our cleaning and sterilization brochure-¿instructions for care, maintenance, cleaning and sterilization of smith and nephew orthopedic devices¿, 74012812.The document is available from customer service or via the smith and nephew website, which suggests using a surgical scrub brush to remove visible debris.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.As the device broke and it cannot longer fit its purpose, the contribution of the device to the reported event could be corroborated.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|