The device was returned for investigation.Upon evaluation of the device, the reported event was confirmed due to faulty power supply.In addition, corrosion on top cover, corrosion on bottom chassis, corrosion on front panel chassis, rear penal deform, rear foot bent, worn out scope socket, cracked lens, and a non-olympus lamp life meter reading over 500+hours were observed.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
|