OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II XENON LIGHT SOURCE Back to Search Results
Model Number CLV-180
Device Problem Optical Problem (3001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported the evis exera ii xenon light source liquid-crystal display fail to light up on the front panel.In addition, the unit makes a clicking noise.The event occurred during reprocessing and there was no patient harm or user injury reported due to the event.This report is being submitted for the failure to light up issue.
 
Manufacturer Narrative
The device was returned for investigation.Upon evaluation of the device, the reported event was confirmed due to faulty power supply.In addition, corrosion on top cover, corrosion on bottom chassis, corrosion on front panel chassis, rear penal deform, rear foot bent, worn out scope socket, cracked lens, and a non-olympus lamp life meter reading over 500+hours were observed.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the faulty power supply could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14027922
MDR Text Key298418152
Report Number8010047-2022-05830
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170215544
UDI-Public04953170215544
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 06/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/11/2022
Initial Date FDA Received04/06/2022
Supplement Dates Manufacturer Received05/26/2022
Supplement Dates FDA Received06/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/28/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-